BO-UP-VF-30 High Purity Water Filtration System Purifier

BO-UP-VF-30 High Purity Water Filtration System Purifier

BO-UP-VF-30 High Purity Water Filtration System PurifierProductFeaturesØApplicable to various water qu
Send Inquiry

 BO-UP-VF-30 High Purity Water Filtration System Purifier

条幅.jpg

Product Features


Ø Applicable to various water quality, the case is with powder coating.

Ø All filter elements in the whole system adopt open mold injection technology

and built-in chassis

Ø Korean quick-connect filter column is easy to replace and maintain.

Ø The electrical and waterways are completely separated to avoid circuit failure

caused by moisture and water leakage.

Ø Water quality online monitoring system.

Ø Automatic RO membrane anti-scaling flushing program, automatic shutdown of tap water, automatic water cut off, automatic water supply in the storage bucket, automatic shutdown of the water storage tank.

Ø All pipelines are NSF certified and have water quality and water upgrade capabilities.

Ø Equipped with a portable TDS pen, you can test TDS content, conductance, water temperature anytime, anywhere.

Ø Double electric water intake, two water quality, ready to use.

Ø Open the door design before and after, it is quick and easy to replace the filter element, and the filter replacement can be easily completed without using any tools.

Product Image

image.pngimage.png

Product parameter

BO-UP series ultrapure water machine

Type classification

Standard

Low Endotoxin

Low TOC

Comprehensive

The machine model

BO-UP

BO-UP-F

BO-UP-V

BO-UP-VF

Water rate

10/20/30 lph

The water requirements

Urban tap water, inlet water pressure: 0.15-0.5mpa, water temperature 5-40℃

RO water Parameter

Ion rejection rate

97%-99% (when using a new RO membrane)

Organic matter rejection rate

>99% when MW>200 Dalton

Particle and bacterial rejection

> 99%

UP water Parameter(ASTM TYPE I grade water)

Resistivity (25℃)

18.25 M Ω. Cm @ 25 ℃

Heavy metal ion

< 0.01 ppb

Total organic carbon (TOC) *

<10 ppb

<3 ppb

bacteria

< 1 cfu/ml *(with terminal filter)

Particulate matter (> 0.1 microns)

<1/ml

Pyrogen (endotoxin)

N/A

< 0.001 Eu/ml

N/A

< 0.001 Eu/ml

Ribonuclease (RNases)

N/A

<1pg/ml

N/A

<1pg/ml

Deoxyribonucleases

N/A

<5pg/ml

N/A

<5pg/ml

Application

All kinds of physical and chemical testing, biochemical analysis, blood examination, microanalysis, flushing special, HPLC, IC, GC, analysis experiment

High sensitivity ICP/MS, PCR, PPT level analysis, isotope analysis, molecular biology, life science, CDC, drug control institute, quality inspection institute

Atomic absorption/emission spectrometry, mass spectrometry, TOC detection, environmental monitoring, biochip development, trace organic matter analysis

Molecular biology, Life sciences, gene research, cell culture, amino acid analysis, protein purification, toxicology, drug development, medical testing

Electrical requirements

100-240V, 50/60Hz, 48-72W

The system configuration

Host machine (including 1 set of purification column)+12L pressure water storage tank +TDS pen + accessory bag

Process(*BO-UP-VF configuration)

image.png

Filter function

image.png

4.jpg