Gas ChromatographyFeatures:★Control system: designed for monitoring and controlling the instrument via the compute
Send Inquiry

Gas Chromatography 

Features:

★Control system: designed for monitoring and controlling the instrument via the computer.

★Column Compartment/oven with superior thermal performance, multistage (10 ramps) programmed temperature control function. (supported by “control system”)

★Advanced built-in data acquisition system , supporting real time instrument status monitoring, detection signal acquisition and PC control

★Column oven accommodates up to 3 chromatographic columns, and supports quick heat-up and rapid cool-down with automated back-door opening.

★Flexible sample introduction system: 3 sample injectors could be installed and operated simultaneously with independent temperature control.

★High sensibility and stability detector.

★2 independent and analog signals output.

★M6 software, compatible with GLP/FDA-21 CFR Part11 requirements and regulations. (electronic records and signatures)


Gas chromatography and accessories:

Carrier Gas System + Sample Introduction System + Separation System + Temperature Control System + Detector

Carrier Gas System: air source/ purification and desiccation device/ flow rate control device

Sample Introduction System: sample injector

Separation System: chromatographic column(packed column and capillary column)

Temperature Control System (Column Oven): constant temperature and programmed temperature

Detector: FID/ TCD/ FPD/ NPD


Advanced Microcomputer Control System:

★Superior performance with advanced, microcomputer-based temperature control system 

★High temperature accuracy (optimum ± 0.02℃), high reliability, and anti-interference

★Self-diagnosis/ self-protection function (overheat protection, power-off protection, etc.)

★Intuitive display of timing program, detector status, measurement range, current setting, etc


Technical parameters:

Model# GC1120
Sample Introduction Systemsample injector and evaporation chamber
Column OvenTemperature rangeAmbient temperature +7℃ ~ 400℃  (in 1℃ increment)
Temperature accuracy ± 0.02℃
Cooling time400℃ to 50℃ in 8-10 min at 25℃ ambient
Programmed temperature setting0.1℃~40℃/min (in 1℃ increment)
Program ramps7 ramps in total (10 ramps available with control workstation)
Size (L*W*H)284*280*241mm(inside)340*345*281mm(outside)
Flame Ionization Detector
(FID)
Detection limit≤3×10-12g/s (C16) 
Best test result≤3×10-12g/s (C16)
Drift≤6×10-13A/30min
Noise≤5×10-14A
Linearity range≥106
Thermal Conductivity Detector
(TCD)
Sensitivity≥5000mV, mL/mg (C16)
Baseline noise≤20µV
Drift≤60µV/h
Linearity range≥104
Flame Photometric Detector
(FPD)
Detection limit≤8×10-13g/s(P) ; ≤8×10-11g/s(S)
Drift≤2×10-11A/30min
Baseline noise≤5×10-12A
Nitrogen-Phosphorus Detector (NPD)Detection limit≤5×10-12g/s(N)  (Azobenzene)
            ≤5×10-13g/s(P)  (Malathion)
Electron Capture Detector(ECD) Linear range≥10 3 (γ -666)
Baseline drift ≤80μ V/h
Baseline noise≤20μ V
Detection limit≤2 X 10-13g/ml (γ -666)